Fluke 52 thermometer is probabaly one of the best built thermometer on the market. It is accurate and built to last. It can read the temperature and the difference of temperature from two thermocouple sensors. One of the features that most other thermocouple based thermometer can't match is stay accurated reading weven when it is subjected to rapid ambient temperature changes (for example, carried form a cold vehicle to a warm buiding). For an ambinet temperautre step change of up to 35C (63F), readings are withing 1 degree of specified accuracy immediately, and within specified accuracy within 10 minutes. It will accept either K type or J type themocouple. The display can be in Celsius or Fahrenheit. User can record the minimum and maximum during a specified period of time.
These thermeters were refurbished from retired equipment of a defence manufacture plant. The back shell of the meter had some drilled holes for attaching to other equipment. They have been pacthed up tith expoxy resine (see picture). Each meter has been reconditioned and calibrated. The meter comes with 90 days warranty.
Sensor is
NOT included. Customer can puchase the K type thermocouple that meet his requirement from this
link. You may also need
thermocouple connector because some of our thermocuople is supplied with spade connector. This thermometer is supplied with a fresh battery.
Click here for Instruction manual
Specifications:
-
Display: Probe 1, Probe 2, Probe 1-2,. Maximum, Minimum, Hold.
-
Display temperature unit in either °C or °F.
-
Input sensor type: K or J type thermocouple
-
Display range: For K-type thermocouple, -200-1370 °C, -328-2498 °F. For J-type thermocouple, -200-760 °C, -328-1400 °F
-
Sampling rate: 1 reading/second
-
Operating condition: Temperature: 0-50°C (32-122°F), Humidity 0-90% RH (0-35C, or 32-95F): 0-80%RH. (35-50C, or 95-122F)
-
Battery type: 9V.
-
Weight: 0.6 lb (including battery).
-
Accuracy:
For K-type +/- 0.1% of reading +0.7 C, or +/- 0.1% of reading +1.3F.
For J-type +/- 0.1% of reading +0.8 C, or +/- 0.1% of reading +1.4F.